Humidity

CAPABILITIES:

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)
Humidity (Source) (10 to 95) % RH 0.59 % RH + 0.058 % of reading
Humidity (Measure and Source) [1] (10 to 95) % RH 1.4 %RH

Infrared

CAPABILITIES:

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)
  Infrared Pyrometers   (35 to 500) °C   0.31 °C + 0.004 3 °C / °C
(95 to 932 °F) 0.56 °F + 0.007 8 °F / °F

Uniformity

CAPABILITIES:

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)
Temperature Uniformity [1] (-90 to 1 000) °C 1.2 °C + 0.059 % of reading
(-130 to 1 832) °F 2.2 °F + 0.059 % of reading
(-90 to 250) °C 0.25 °C + 0.015 % of reading
(-130 to 482) °F 0.45 °F + 0.015 % of reading

Thermometers & Probes

CAPABILITIES:

Parameter/Equipment Range Expanded Uncertainty of Measurement (+/-)
Temperature Measure [1] (-196 to -100)°C 0.04 °C
(-321 to -148) °F 0.07 °F
(-100 to 0)°C 0.012 °C + 0.065 % of reading
(-148 to 32) °F 0.019 °F + 0.065 % of reading
(0 to 660)°C 0.012 °C + 0.000 7 % of reading
(32 to 1220) °F 0.022 °F + 0.000 7 % of reading
Temperature Measuring Equipment [1] -196 ±5 °C 0.04 °C
-321 ± 9 °F 0.07 °F
(-100 to 70) °C 0.012 °C + 0.006 5 % of reading
(-148 to 158) °F 0.019 °F + 0.006 5 % of reading
 (60 to 300)°C 0.012 °C + 0.001 1 % of reading
(140 to 572) °F 0.022 °F + 0.001 1 % of reading
(150 to 400) °C 0.012 °C + 0.001 1 % of reading

(302 to 770) °F

 

0.022 °F + 0.001 1 % of reading

 

ITS 90 – Fixed Point 660 °C 0.54 °C

1220 °F

 

0.97 °F

 

Surface Temperature Measurement [1] (0 to 250) °C 1.3 °C
(32 to 482) °F 2.3 °F

If you would like a quote, feel free to fill out our Contact Form.